Introduction To Splitters

Introduction

Early microscopes were essentially a tube through which light travels (Figure 1A), from a sample to the eye (or a camera), through some lenses. Modern microscopes have a variety of objectives, mirrors, and pinholes in order to obtain the best image (Figure 1B). The component of interest here is the beam splitter.

Figure 1: The light path through different microscopes. A) An early compound microscope with a basic light path. The light goes from the object, through the objective, tube, and eyepiece, into the eye or a camera. B) A modern confocal microscope, where light comes in from the side, is redirected by the beam splitter through the microscope objective and through the sample. Returning light from the sample goes through the same objective and beam splitter, through a pinhole and into a detector (typically a scientific camera).

A beam splitter is an optical device that splits beams (such as laser beams) into two (or more) beams. Beam splitters typically come in the form of a reflective device that can split beams into exactly 50/50, half of the beam being transmitted through the splitter and half being reflected. Fig.2 outlines the uses of beam splitters.

Figure 2: The use of beam splitters. A) A cube beam splitter. 1 is the incident (entering) light beam, 2 is 50% transmitted light, 3 is 50% reflected light. This may not always be exactly 50/50 as the reflective part can absorb some light. B) Some cube beam splitter cubes with ruler for scale. C) A cube beam splitter with the incident light on the bottom left and the two split beams on the right. D) A mirror beam splitter splitting a green laser.

Common types of beam splitter are either cube beam splitters or plate beam splitters (such as mirrors), as described below.

Cube beam splitters are made from two triangular glass prisms glued together, as seen in Fig.2A-C. The thickness of the glue is carefully adjusted so that, for certain wavelengths of light, half the light that enters the cube is reflected 90° and the other half is transmitted through the cube in the same direction as it entered, as seen in Fig.2A.

Half-silvered mirrors are a sheet of glass or plastic with a very thin coating of reflective metal, usually aluminum. The thickness of the metal coating allows for half the light to be transmitted, and the other half to be reflected at 90°. An example of this mirror can be seen in Fig.1D.

Dichroic mirrors are similar to metallic-coated mirrors but instead use a dichroic optical coating. Dichroic materials cause light to be split into distinct beams of different wavelengths (from the Greek dikhroos, meaning two-colored), depending on the material used the ratio of reflection/transmission can be changed.

Dichroic mirrored prisms are prisms that use a dichroic optical coating and can split beams up to three times. These devices could also be used in reverse, as a beam combiner.

When comparing plate/mirror and cube beam splitters, the mirror splitters can tolerate more powerful beams of light, but the cubes have far better durability and are easier to handle. While both mirror and cube beam splitters can be used for simple light beams, they can also split beams carrying an image, which makes beam splitters a powerful tool for microscopy.

Splitter Applications

Two major examples of the usefulness of beam splitters are:

Emission image splitters: splitting image light from a microscope and re-aligning it so one camera can detect multiple images (Fig.3A)

Multiple camera adapter: splitting image light from a microscope into multiple cameras so that each one handles a specific wavelength (Fig.3B)

Figure 3: Two microscopy devices that utilize beam splitters for microscope applications. A) An emission image splitter, where image light from the microscope is split, reflected and carefully combined at an offset. This means that each half of the camera sensor receives each signal, so one camera can image two different wavelengths (with the sensor split in half). B) A multiple camera adapter, where the image light is split to two separate cameras. Each camera can image a specific wavelength of the same sample.

Emission Image Splitter

Emission image splitters allow for a single camera to image at multiple wavelengths by splitting the camera sensor into sections and projecting emitted light onto each part of the sensor. The optics within an emission image splitter can be seen in Fig.4.

Figure 4: Internal workings of an emission image splitter. Light travels from left to right in this diagram (as indicated by the big arrow). Emitted light from a sample enters the splitter from the microscope and is split into two channels based on wavelength. These channels are manipulated by mirrors to the same camera sensor but offset so that each channel occupies one half of the camera sensor. The sensor is split vertically into two halves, and two images are observed from one sample.

These splitters act as an interface between the microscope and the camera, emitted light from the sample passes from the microscope to the splitter, and are split based on wavelength before being projected onto sections of the camera sensor. Splitters can split images two, three or even four times based on wavelengths, allowing researchers to image multiple fluorophores simultaneously rather than having to switch channels manually or electronically. Examples of emission image splitters can be seen in Fig.5.

These splitters all attach to standard C-mount ports on microscopes and offer standard output ports for cameras. This makes these splitters a seamless interface, allowing for multiple images on a single camera.

Emission image splitters have a wide variety of uses, as they can split an image across a camera sensor based solely on wavelength, polarization or amplitude. The main advantage is the simultaneous imaging at multiple wavelengths. If researchers want to image two different fluorophores, typical imaging systems involve manually or electronically cycling through filters in order to image at the desired wavelength. With a splitter, both wavelengths are imaged simultaneously, suitable for long term experimentation, fast dynamic events and any imaging setup that involves multiple fluorescent probes. Brightfield and fluorescence can be imaged simultaneously if desired, and splitting based on polarization allows for flexible experimental setups. Advanced microscopy techniques such as voltage/calcium imaging, Förster Resonance Energy Transfer (FRET), spinning disk confocal and Total Internal Reflection Fluorescence (TIRF) could all benefit from an emission image splitter.

Figure 5: Different formats of emission image splitters. Each splitter acts as an interface between the microscope and the camera, splitting an image into two, three or four based on wavelength, as shown by the color cube. The three emission image splitter models shown here are the OptoSplit II (two-way split), OptoSplit III (three-way split) and the MultiSplit V2 (four-way split). The two- and four-way splitters are shown installed on a microscope system, while the three-way splitter shows an example of a triple split image.

While an emission image splitter can greatly enhance any imaging system by allowing for simultaneous imaging at 2/3/4 different wavelengths, the main disadvantage is that each channel occupies space on the camera sensor. When using a two-way splitter, the camera sensor is cut in half in order to resolve two images simultaneously. The two images have half the resolution and field of view due to only having access to half a sensor each. With large camera sensors, this isn’t such an issue, but if a four-way splitter is used on a smaller camera sensor, images may not be resolved properly due to only having a small portion of the sensor to interact with.

Multiple Camera Adapter

While an emission image splitter allows for multiple images on a single camera, the multiple camera adapter does the opposite: allows multiple cameras to image the same sample. As seen in Fig.3B, a single splitter sends half the light (reflected) from the microscope to one camera, and the other half (transmitted) to a second camera, split based on wavelength, polarization or amplitude. This setup allows for multiple cameras on one microscope port.

Different multiple camera adapters allow for increasing numbers of cameras on the same port, with up to four cameras imaging the same sample. These devices can be seen in Fig.6.

Figure 6: Multiple camera adapters. The top splitter is the TwinCam, using a single mirror splitter to allow up to two cameras on one microscope port. The bottom splitter is the MultiCam, using two mirror splitters to allow up to four cameras on one microscope port. These multiple cameras can simultaneously image the same sample.

These splitters work as an interface between the microscope and the cameras, opening up a single microscope port for use with as many cameras as desired. While using an emission image splitter involves decreasing the camera sensor size for each image in order to acquire images simultaneously, there is no such trade-off with the multiple camera adapter, both cameras retain their full field of view and resolution but can each image the same sample at a different wavelength. This allows for the full potential of each camera to be utilized. In addition, if imaging at challenging wavelengths (such as ultraviolet (UV) or near-infrared (IR)), one camera can be used to cover these wavelengths, while still imaging at conventional fluorescence wavelengths with another camera.

The main drawback is the need to buy multiple cameras as well as a splitter, making the multiple camera adapter the more expensive option. Despite this additional cost, the significant increase in resolution and field of view make the multi-camera adapters an attractive option for simultaneous imaging.

Summary

In summary, if your experiments would benefit from simultaneously imaging your sample at different wavelengths, polarization states or amplitudes, splitters allow for more flexible experimentation. Whether this involves one image being split across one camera or one sample being split across multiple cameras, the numerous splitter devices available allow for more customizable research and better multi-channel imaging.

For more information about splitters, please refer to our other tech notes on this subject.